To clean the probe:
1. Inspect the probe's lens, cable, casing, and connector for cracks, cuts, tears, and other signs of physical damage.
2. Disconnect the probe from the ultrasound console and remove all coupling gel from the probe by wiping with a soft cloth and rinsing with flowing water.
NOTE: DO NOT wipe the probe with a dry cloth.
3. Soak the probe head in water. Scrub the probe as needed using a soft sponge, gauze, or cloth to remove all visible residue from the probe surface.
4. Rinse the probe with enough clean potable water.
5. Air dry or dry with a soft cloth.
6. After cleaning, inspect the probe's lens, cable, casing and connector. Look for any damage that would allow liquid to enter the probe. Also, inspect the probe functionality by live scan. If any damage is found, do not use the probe until it has been inspected and repaired/replaced by a GE service representative.
NOTE: Do not immerse the probe into any liquid beyond the level specified for that probe. Never immerse the transducer connector into any liquid.To disinfect a probe:
1. After cleaning, the probe and cable may be wiped with a tissue sprayed with a recommended disinfectant.
2. After disinfecting, inspect the probe's lens, cable, casing and connector. Look for any damage that would allow liquid to enter the probe. Also, inspect the probe functionality by live scan. If any damage is found, do not use the probe until it has been inspected and repaired/replaced by a GE service representative. Refer to the Probe Care Card enclosed in the probe case or to www.gehealthcare.com/transducers for the latest list of compatible cleaning solutions and disinfectants. -